Warning:
Proper functioning of the disposable electrodes is influenced by large amounts of
body hair, cream or oily skin. Regularly remove hair from your skin at the attachment points of
the electrodes and ensure that your skin is free of cream and fat. Do not clean your skin with
alcohol.
Warning:
Do not record a reference ECG with incorrectly positioned or transposed electrodes,
as a correct evaluation of subsequent control ECGs is not possible.
Warning:
Renew your reference ECG after every acute event, intervention (e.g. heart attack,
cardioversion in case of arrhythmia, heart bypass intervention, heart pacemaker implantation,
cardiac catheter intervention, heart valve intervention), medication change or when recom-
mended by your physician.
Warning:
Only use the disposable electrodes recommended by Personal MedSystems (YD43-6-
W or P-00-S/xx).
Warning:
Only use disposable electrodes that are unused, have not expired and are at room
temperature, in order to ensure sufficient signal quality and prevent distorted results.
Warning:
Remove all disposable electrodes after performing a reading, as prolonged adhesion
may cause skin irritation.
Warning:
Do not use CardioSecur in case of sensitive skin or allergies. This could cause rash,
irritation of the skin or eczema.
Warning:
Only use the ECG cable as described in this user manual and provided by Personal
MedSystems GmbH.
36
V7.1