3
APPLICATION
3.1 Connecting the pressure gauge
• Fit the test nozzle onto the appropriate connector on the
pressure gauge.
• Fit the connection tubing onto the appropriate connector on
the pressure gauge.
• Connect the other end of the connection tubing to the air
connection on the compressor.
Depending on the compressor, the tubing adapter may need
to be used.
3.2 Checking the operating pressure
Check the operating pressure of PARI compressors once a
year. For rented devices, check the operating pressure every
time they are rented. This is the only way to ensure it is in good
working order.
• Switch on the compressor and leave it running for at least
10 seconds.
• Read the operating pressure measured on the pressure
gauge.
• If the operating pressure is less than the values stated in the
table below, contact PARI Service.
– 7 –