Application; Connecting The Pressure Gauge; Checking The Operating Pressure - Pari PG 041 Notice D'utilisation

Appareil de contrôle
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3

APPLICATION

3.1 Connecting the pressure gauge

• Fit the test nozzle onto the appropriate connector on the
pressure gauge.
• Fit the connection tubing onto the appropriate connector on
the pressure gauge.
• Connect the other end of the connection tubing to the air
connection on the compressor.
Depending on the compressor, the tubing adapter may need
to be used.

3.2 Checking the operating pressure

Check the operating pressure of PARI compressors once a
year. For rented devices, check the operating pressure every
time they are rented. This is the only way to ensure it is in good
working order.
• Switch on the compressor and leave it running for at least
10 seconds.
• Read the operating pressure measured on the pressure
gauge.
• If the operating pressure is less than the values stated in the
table below, contact PARI Service.
– 7 –

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

007

Table des Matières