6.
Remove toddler from the scale and then record weight. The weight will remain on
the display for two minutes.
7.
Although the BabyChecker will automatically turn off after about two minutes, slide
the lb•oz/off/ kg switch to OFF. Then if the Start/Zero button is accidentally
pressed, the batteries will not be used.
TIPS FOR USING THE BABYCHECKER
When pressing the Start/Zero Button, do not touch any other part of the scale.
Touching other parts might cause the scale to display negative figures such as –0.04
or –0.10, therefore making the accuracy of the weight calculation invalid.
When lifting or moving the scale, be sure to lift the entire scale from the bottom of
the Scale Housing, not the Baby Basket.
Do not put water on the scale or place the scale in a humid location.
The scale is a precision instrument. Do not drop it, jump on it, or drop a heavy
object onto its surface.
Do not dismantle the scale.
CARE AND CLEANING INSTRUCTIONS
The scale is a precision instrument and specially adjusted and manufactured for accu-
rate measuring. Use of the scale following these instructions is important.
1.
Do not dismantle the Scale Housing.
2.
The scale should not be operated in the presence of external vibration or direct
sunshine.
3.
Rapid change of ambient temperature may affect the accuracy of the scale for up to
2 hours.
4.
Do not spill water or chemical liquid on or in the Scale Housing. Periodically, wipe
the scale with a damp cloth. The Baby Basket should be cleaned using Cavicide
another non–staining germicidal cleaner.
5.
Keep the scale away from any heating appliances (oven, stoves, etc.).
6.
Keep the unit in a cool, dustless and dry environment and remove batteries during
extended storage.
7.
When in storage, do not put anything on the Baby Basket.
All manuals and user guides at all-guides.com