Operating Messages; Error Codes; Battery Disposal - Xylem 910 COD Mode D'emploi

Colorimeter
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operating Messages & error Codes

Operating messages

Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Lo
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
btLo
measurement is still possible.
A user calibrated method is indicated by an arrow while the test result is
Store
Date
Cal
RESULT
displayed (see "Factory calibration reset").
Time
Cal

error codes

Light absorption too great. Reasons: e.g. dirty optics.
E27 / E28 / E29
Calibration factor „out of range"
E 10 / E 11
too much light reaching the detector.
E 20 / E 21
E23 / E24 / E25
too much light reaching the detector.
Battery capacity was too low during measurement. Change battery.
E 22
LR: Factory calibration incorrect / erased
E 70
LR: User calibration incorrect / erased
E 71
MR: Factory calibration incorrect / erased
E 72
MR: User calibration incorrect / erased
E 73

Battery Disposal

The 910 COD Colorimeter is powered by alkaline batteries which the user must
remove and dispose of when the batteries no longer power the instrument. Dis-
posal requirements vary by country and region, and users are expected to under-
stand and follow the battery disposal requirements for their specific locale.
englIsH
13

Publicité

Table des Matières
loading

Table des Matières