SERVICE

CAUTION
To assure product safety and reliability, repairs should be performed by
Magnaflux Authorized Service Center using original parts.
WARNING
The LED assembly is static sensitive and can be permanently damaged with
improper handling. All replacements of the LED assembly must be performed by a
Magnaflux Authorized Service Center. Unauthorized repairs will void the warranty.
WARNING!
To maintain cTUVus certification, service of model 628785 must be performed by
the factory. Contact Magnaflux Customer Service for more information.
UV FILTER/GLASS REPLACEMENT
1.
Disconnect the lamp from the power supply.
2.
Remove the protective shade from the front of the lamp.
3.
Using a 3-mm hex key, remove the four screws from the front bezel.
4.
Remove the front bezel, then remove the UV filter/glass and gasket.
5.
Inspect the gasket for tears or damage. If damage is found, replace the gasket.
6.
Install the new filter/glass, placing the gasket below the filter/glass in the plastic inset tray.
7.
Replace the front bezel and secure with the four hex screws. NOTE: Tighten hex screws
until they stop (little to no force required), then tighten an additional 1/8 turn.
8.
Replace the protective shade.
CORD ASSEMBLY REPLACEMENT
1.
Disconnect the lamp from the power supply.
2.
Using a 3-mm hex key, remove the four bolts from the rear cover.
3.
Remove the rear cover (13) and gasket (10). Disconnect the blue leads (11) from the switch assembly (12).
4.
Remove the 4-pin latch connector (8) from the LED assembly (7). Note the wire
configuration used on the connector for installation of the new cord.
5.
Using wire cutters, cut the brown and gray wires to remove the 4-pin latch connector from the cord.
6.
Remove the ground wire (green/yellow) from the inner top of the housing.
7.
Using a 16-mm wrench, loosen the nut on the cord grip (16) at the bottom of the handle.
8.
Use the wrench to remove the entire cord grip from the handle. Slowly pull the old cord
out of the handle.
9.
Remove the nut from the new cord grip included in the replacement kit. Use the wrench to
install the new cord grip in the handle.
10.
Slip the new cord grip nut onto the new cord. Insert the new cord through the cord grip in
the handle and carefully work the cord through to the lamp housing.
11.
Attach the ground wire (green/yellow) to the top of the housing and tighten in place.
12.
Using the wire configuration noted in Step 4, insert the gray and brown wires into the new
4-pin latch connector included in the replacement kit.
13.
Attach the 4-pin latch connector to the LED assembly.
14.
Using the wrench, tighten the cord grip nut at the bottom of the handle to secure the new
cord in place.
15.
Reconnect the blue leads to the switch assembly on the rear cover.
16.
Replace the rear cover and secure with the four bolts. Note: tighten bolts until they stop
(little to no force required), then tighten an additional 1/8 turn. Do not over-torque.