FaCtOrY CaliBratiOn reset
Resetting the user calibration to the original factory calibration will reset all meth-
ods and ranges.
A user calibrated method is indicated by a "Cal" symbol while the test result is
Store
Date
Print
Time
displayed.
Cal
RESULT
To reset the calibration press both the [MODE] and [ZERO/TEST] key
Zero
Mode
and hold.
Test
Turn the unit on using the [ON/OFF] key.
On
Off
Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.
The following messages will appear will flash on the display:
SEL
The factory setting is active.
CAL
(SEL stands for Select)
or:
SEL
Calibration has been set by the user.
(If the user calibration is to be retained, turn the unit off using the
cAL
[ON/OFF] key).
Calibration is reset to the factory setting by pressing the [MODE] key.
Mode
The following messages will appear in turn on the display:
SEL
CAL for factory calibration; cAL for user calibration.
CAL
On
Turn the unit off using the [ON/OFF] key.
Off
ENGLISH
19