BRAVO
Operator's Manual (located on the USB key) should be consulted before using this autoclave.
BASIC OPERATION
1
Power unit ON.
A
3
Load configurations.
4
Close unit door.
6
Open door.
A
MAINTENANCE SCHEDULE
Clean the door gasket and internal part of the door
with a clean wet cloth. Check for wear and replace
if required.
Daily
Completely empty the waste tank before refilling
the water reservoir.
Clean the chamber and appropriate accessories
with a wet cloth. If required, use a neutral detergent
(without alkalines and bleaches). Carefully rinse with
demineralized/distilled water and remove any residue.
Weekly
Inspect and clean chamber anti-dust filter.
Replace if necessary.
Ensure area around BRAVO G4 is dust free.
Drain & clean the internal water reservoir as per
the instructions outlined in the Operator's Manual.
Monthly
If applicable, clean the external water tank used for
automatic filling.
www.scican.com
97055284-1 | SD-622 NA EN R1. BRAVO is a trademark of SciCan Ltd. A Coltene Group Company.
Manufactured for SciCan Ltd. 1440 Don Mills Rd. Toronto, ON, M3B 3P9 / Canada
G4
™
Quick Reference Guide
2
Fill with demineralized/distilled water.
5
Run cycle.
47'
132°C Hollow wrapped B
Start a cycle.
Select a cycle.
7
Unload instruments.
Every
3 - 6
Months
Every
Year
or
1,000
Cycles
NEVER USE
TAP WATER
132°C Hollow wrapped B
132°C
1,90 bar
132°C Hollow wrapped B
CYCLE
COMPLETED
133,3°C
133,1°C
1,941 bar
1,929 bar
End of cycle.
8
If required, drain manually.
Check the bacteriological filter and replace
as required.
Lubricate the door lock. Using a clean cloth, remove
any residues and lubricate the inside of the bushing
with a film of the silicone-based grease provided.
Inspect and clean
the chamber
drain filter.
Replace if required.
Replace the door gasket.
+1 800-572-1211
+1 800-667-7733
TOUCHSCREEN
Data
Settings
VENTILATION
Front & Side
Back
0.78in / 20mm
0.78in
0.78in
20mm
20mm
FILLING OPTIONS
Top
Filling
Front
Filling
DRAINING OPTIONS
Manually
Automatic
To Release
Spare Parts and Accessories
97400145
Door Gasket 17L/22L
97290160
Bacteriological Filter
97290210
Chamber Drain Filter
99951785
Chamber Rack 17L
99951786
Chamber Rack 22L
99950394
Pouch Rack
99951317
Tray 17L
99951318
Tray 22L
97467066
Tray Extractor
IMPORTANT!
Users of this device must be sure to select
the most appropriate sterilization cycle
according to the recommendations of the
leading infection control authorities and local
regulatory guidelines / recommendations.
Cycles
Solid
Solid, Non-Lumened
Instruments
Hollow
Hollow, Lumened
Instruments
Rubber & Plastics
Textiles, Hollow,
Lumened, Solid