Bolus Volume Accuracy; Spare Parts - IVAC P5000 Mk II Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

The following table provides an indication of the accuracy with which a bolus infusion will be
delivered. Test carried out as specified in IEC601-2
Bolus Volume
Bolus Rate
(ml)
(ml/hr)
0.1
100
2.0
100
5.0
100
A comprehensive list of spare parts for IVAC Syringe pumps is included within the service manual.
This can be ordered from ALARIS Medical Systems, or authorised distributor. For part number
please refer to summary parts list below:
Part Number
5001FAOPT71
5000PB0004
5000SP00008
5000SP00010
1000SP01008
0000EL00004
0000EL00287
0000EL00280
1001FAOPT91
1001FAOPT92
1001FAOPT93
0000ME00026
1000SP01015
5000LB00020
5000LB000023
5001FAOPT71 ISS 8.0
No. of Samples
Max. Positive
(%)
25
+12.0
25
+2.5
25
+1.0
Description
PCAM Directions for Use P5000 Mk II GB/FR/D.
PCAM Service Manual. Model P5000.
Printer Cable 9 to 25 pin (9 - pin female (PCAM) to 25 - pin
male (printer)).
Citizen PN60 Printer Cable (9 - pin female (PCAM) to 26 - pin
AMP17823404).
Comms Cable (9 - pin female to 9 - pin female)
Internal Battery - 6v NP2.6-6 Rechargeable.
Fuse - T63mA (Time Lag 5 x 20mm) - 220/240V units
Fuse - T12mA (Time Lag 5 x 20mm) - 110/120V units
AC Power Cable - UK
AC Power Cable - European
AC Power Cable - Unterminated
Foot - Self Adhesive
Pole Clamp Assembly
Label Set P5000
Label - Front Panel
Specifications

Bolus Volume Accuracy

Max. Negative
(%)
-14.0
-0.0
-0.0

Spare Parts

35 / 120
Mean
(ml)
-5.0
+1.0
+0.8