6 Limitations on Use:

Intended Life:
A local risk assessment should be carried out based upon activity and usage.
Lifting Loads:
User weight and activity is governed by the stated limits.
Load carrying by the user should be based on a local risk assessment.
Environment:
The product is water proof to a maximum depth of 1 metre.
Thoroughly rinse with fresh water after use in abrasive environments such as those that may
contain sand or grit, for example, to prevent wear or damage.
Thoroughly rinse with fresh water after use in salt or chlorinated water.
Exclusively for use between -15˚C and 50˚C (5˚F to 122˚F).
It is recommended that only Endolite products be used in conjunction with the product.
7 Heel Spring Removal & Replacement
1
10
Undo the 2 Heel Spring retaining bolts.
2
10 10Nm
Fit the Toe Spring into the Heel Spring cover and
tighten the bolts to 10Nm.
Replace the Heel Wedge if fitted.
6
938371/03-1016