Calex PyroMini Serie Guide De L'opérateur page 6

Publicité

Les langues disponibles
  • FR

Les langues disponibles

USer interface
default View
setting
Temperature
units
selecting
displayed
Temperature
6
Temperature View
Displays a large indication of the measured temperature. The background
turns bright red when an alarm is activated.
Press "°C" to switch to °F and vice versa. The units are changed throughout
the interface.
Press the temperature display to select which reading is shown:
filtered Temp
The measured temperature, with averaging and hold processing. This
temperature is output by the sensor on the 4 to 20 mA output (-CB and -CRT
models).
average Temp
The measured temperature with averaging but without hold processing.
Unfiltered Temp
The unprocessed measured temperature.
MicroSD Card status.
This icon is displayed when an SD card is inserted, and flashes when data
logging is in progress.
This icon is displayed when scheduled data logging is enabled and has yet
to begin.
list View
Displays a list of the measured temperatures, alarm state and data logging
state.
filtered Temp: The measured temperature, with averaging and hold
processing.
Unfiltered Temp: The unprocessed measured temperature.
average Temp: The Unfiltered Temperature averaged over the period
specified in "Output Processing".
Maximum Temp: The highest temperature measured during the hold period,
with averaging.
Minimum Temp: The lowest temperature measured during the hold period,
with averaging.
sensor Temp: The internal temperature of the sensing head.
Reflected Temp: The reflected energy compensation temperature, as
specified in "Emissivity and Compensation".