Télécharger Imprimer la page

Calex PyroMini Serie Guide De L'opérateur page 10

Publicité

Les langues disponibles
  • FR

Les langues disponibles

SettingS > alarmS
alarm set
point
hysteresis
filtered
Temperature
or sensor
Temperature
(alarm 2 only)
alarm Type
reset
log Trigger
Time
log data
While
Triggered
log
acknowledge
Time
log reset
Time
1 0
alarm 1 and alarm 2
The temperature at which the alarm is triggered. Minimum: -20°C. Maximum:
1000°C.
The temperature difference between the Alarm Set Point and the reset
temperature. Hysteresis is only used when Automatic Reset is enabled.
Please see the diagrams below for more information
Minimum: 0°C (hysteresis disabled). Maximum: 1000°C.
Select the temperature monitored by Alarm 2.
high
The alarm is triggered when the temperature rises above the Alarm Set Point.
low
The alarm is triggered when the temperature drops below the Alarm Set
Point.
Off
The alarm is disabled.
automatic
The alarm is acknowledged and reset automatically when the temperature
reaches the reset temperature (see Hysteresis). It can also be acknowledged
or reset manually.
Manual
The alarm is acknowledged by pressing Acknowledge
Temperature View or List View, and reset by pressing Reset
Alarms screen.
alarm logging
Alarm events can be logged to the MicroSD Card. Alarm log files and settings
are independent from Data Logging.
The time that an alarm is triggered will be logged.
Data logging will start when an alarm is triggered. 1 sample is logged per
second. Logging stops when both alarms are reset.
The time that the alarm is acknowledged will be logged.
The time that the alarm is reset will be logged.
SEN TRONIC
056 222 38 18
AG
www.sentronic.com
mailbox@sentronic.com
on the
on the

Publicité

loading