E L PA RTS
Operating manual
SelectH
selective gas detector
2
5.1
Error Codes
ERR P
Problem with the gas flow
<
E 32
Calibration data error
<
E 64
Sensor break
<
E 128
Parameter store faulty
<
E 192
Parameter store faulty and sensor break
<
6
Charging the batteries
The device must be switched off during charging. The mains adapter
provided (max. charging current 150 mA) is used to charge the batte-
ry within 12 to 14 hours. The red LED on the back of the housing sig-
nalises the charging process and is permanently lit during charging.
The charging procedure should take place in the temperature range
of 0 ... 45°C.
Under normal conditions, more than 500 charging and discharging
cycles can be expected. Long-term storage or excessively long batte-
ry charging leads to a reduced cell capacity. This loss of capacity can
be compensated by a full charge or by 1 to 3 charging and dischar-
ging cycles.
7
Device-related requirements according to
DVGW G 465-4 and technical data
7.1
General device-related requirement
according to DVGW G 465-4
This DVGW reference concerns mobile devices used for detecting
leak spots on public gas supply systems, as well as on the customer's
own system, (domestic installations).
The term "gas detection devices" is used to describe devices which
prove gas concentrations with a sensitivity level of <100 ppm fuel
gas, in other words devices that are usually used to prove the exis-
tence of gas during leak detection.
The properties required of the devices should – as far as is appro-
priate – be listed in the manufacturer's operating instructions.
7.2 Technical data
Measuring range:
Resolution:
Measuring components / testing gas:
Functional principle:
Measuring result display:
through diodes, measuring values in ppm and
Time to stand-by:
Reaction time (T90):
Permissible working temperature:
Permissible storage and transport temperature:
Permissible storage and transport humidity:
Signalbehaviour when measuring range is exceeded:
met, lower explosion limit display flashes
Automatic zero point setting:
Acoustic signal:
permanent warning signal when the lower explosion
Battery state:
diode flashes, remaining operating time about 15 min
Fault message:
Dimensions:
180 x 50 x 28 mm (length x width x height)
Weight:
Power consumption:
Operating time with battery fully charged:
Battery:
4 x 1.2 V 1600mAh, to be charged with 9 V
Mains adapter:
Input voltage:
Output voltage:
16
H
: 1ppm – 999ppm
2
0,1 Vol% - 4,0 Vol% / UEG
1ppm / 0,1 Vol%
1 Vol% hydrogen
in synthetic air
Gas-sensitive semiconductor
GGS 1000/6000
Threshold display
from 0 ppm to 1.0 Vol.%
Vol.% on LC display up to UEG
< 50 seconds
2-3 seconds
-15 ... +40°C
-25 ... +70°C
20 ... 80% %
rel. humidity (non-condensing)
Demand
given at > 999 ppm
and > lower explosion limit
Controlled by concentration,
can be switched off,
limit is exceeded
Battery symbol in display, „operation"
See list of fault codes on page 14
410 g in total
at least 4 hours
230V~ / 50Hz / 3.5VA
9V = / 150mA / 1.5VA
2,00 VA