OxyGo FIT® Column Change Procedure (continued)
9. Insert column assembly into the OxyGo FIT® con-
centrator. Do not leave the column ends exposed;
column assembly should be inserted into the OxyGo
FIT as soon as the dust caps have been removed.
10. Push the column assembly into the device such that
the columns are fully seated into the OxyGo FIT®
concentrator. The spring loaded latch button should
fully return to the closed position.
11. Connect the AC power supply cord to the OxyGo FIT®
and plug the power supply AC cord into an
electrical outlet. Do Not Power on the OxyGo FIT®
concentrator.
12. Press and hold the plus (+) and (-) minus buttons for
5 seconds. The screen will display "sieve reset"
message. Release button once message is displayed
on screen.
13. Press the alert
display "sieve reset success".
14. Press the power
FIT®, and use normally.
WARNING Do not use any columns other than those specified in this user manual.
C
!
96-07062-00-01B OxyGo FIT Multilingual Manual.indb 35
button once and screen will
button to turn on the OxyGo
The use of non-specified columns may create a safety hazard and/or
impair equipment performance and will void your warranty.