4. Install a replacement fuse, OxyGo FIT® RP#125 (BUSS MDA-12) and reassemble the tip.
Ensure the retainer ring is properly seated and tightened.
Cigarette Adapter Plug
CAUTION
For continued protection against risk of fire, only use fuse specified.
OxyGo FIT® Column Change Procedure
NOTE
Column change instructions are only to be used when maintenance is
required and are not intended for practice purpose.
1. Turn off the OxyGo FIT® concentrator by pressing the power button to shut down the
device.
2. Remove the OxyGo FIT® concentrator from the carry case, if applicable..
3. Remove the battery from the OxyGo FIT® concentrator.
4. Place the OxyGo FIT® concentrator on its side so that
the underside is visible. The metal column assembly
can be seen on one side of the device.
96-07062-00-01B OxyGo FIT Multilingual Manual.indb 33
Retainer Tip
Fuse
Cleaning, Care and Maintenance Chapter 6
Retainer
33
11/16/2016 1:09:41 PM