1. YOUR NEW MICROPIPETTE
Your new hand held micropipette EASY 40 Elite is a general purpose pipette for the
accurate and precise sampling and dispensing of liquid volumes. The micropipettes
operate on the air displacement principle and disposable tips.
The micropipettes cover a volume range from 0.1µl to 5ml.
All micropipettes have been quality tested according to ISO8655/DIN12650. The
quality control according to ISO8655/DIN12650 involves gravimetric testing of each
pipette with distilled water (quality 3,
DIN ISO 3696) at 22ºC using the manufacturer's original tips.
1.1. Adjustable volume pipettes
Volume Range
0.1 -2.5µl
0.5-10µl
2-20μl
5-50μl
10-100μl
20-200μl
100-1000μl
1000-5000μl
1.2 Fully autoclavable
The micropipette can be fully autoclaved, withstanding steam sterilizing at 121ºC,
1 atm. for 20 minutes. Single channel micropipettes can be autoclaved without
special preparation. After autoclaving the micropipette must be cooled down and
left to dry over 12 hours. It is recommended to check the performance of the
micropipette after each
and seal of the pipette after 10 autoclavings.
2. UNPACKING
The micropipette packing contains the following items:
Micropipette
Calibration/ remove tool
User manual
Micropipette holder
Sample Tips
Quality control certificate
3. INSTALLING THE MICROPIPETTE HOLDER
For convenience and safety always keep the micropipette vertically on its own
holder when not in use. When installing the holder, please follow the instruction
below:
1. Clean the shelf surface with ethanol.
2. Remove the protective paper from the adhesive tape.
3. Install the holder as described in Figure 2A. (Make sure the holder is pressed
against the edge of the shelf.)
4. Place the micropipette onto the holder as shown in Figure 2B.
Fig. 2A
2
User Manual
Variable volume micropipette
EASY 40 Elite
Increment
0.05µl
0.1µl
0.5μl
0.5μl
1μl
1μl
1μl
50μl
autoclaving.
It is also recommended to grease the piston
Recommended Tips
10µl
10µl
200, 300μl
200, 300, 350μl
200, 300, 350μl
200, 300, 350μl
1000μl
5ml
Fig. 2B