Warnings/ Caution - Cuda surgical Elite 1000 Manuel De L'opérateur

Masquer les pouces Voir aussi pour Elite 1000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

2. WARNINGS/ CAUTION

WARNING
The illuminator uses a highly concentrated light source (luminous power per
area) and this high energy density is retained through connected lightguides and instruments.
The output of a connected instrument left in close proximity or contact with tissue or flammable
materials presents a risk of injury or fire. Qualified personnel must determine a safe working
distance and intensity setting for each application. The output should never be left on
unattended.
Caution Rx only. Federal law restricts this device to sale by or on the order of a licensed
healthcare practitioner.
Caution To prevent fire or electric shock, do not open or expose the lightsource unit to
rain or moisture. Refer all servicing to qualified personnel only.
Caution Not suitable for use in presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
Caution To prevent any potential electro-magnetic interference, do not use any kind of
cellular phone near the lightsource.
Caution This product should be used only with type BF endoscopic instruments which
have been certified according to IEC 60601-1 for medical equipment and IEC 60601-2-18 for
endoscopic equipment.
Caution User must not alter this device in any fashion. Doing so voids all warranties and
statements of suitability for any purpose.
Caution
All devices connecting to the Illuminator must be classified as medical
equipment. Additional information processing equipment connected to the Illuminator, a Medical
System and the operator must determine that all equipment complies with the appropriate end-
product standards (such as IEC 60950 or IEC 60065 and the Standard for Medical System, IEC
60601-1-1).
Caution Always set the intensity control to the minimum level and insert the fiberoptic
cable into the unit before turning on the power. When light is not required at the surgical site,
the intensity control should be should be set to the fully dimmed position.
necessary to remove the fiberoptic cable without turning the unit off, turn the intensity control to
the fully dimmed position.
Caution The fiberoptic cable must be electrically NON-CONDUCTIVE. It should not
have conductive shielding or any other conductive connection between the patient and
equipment. Such connection will impair safety of the equipment. It must be rinsed free of
soaking/disinfectant solution and dried before plugging into the lightsource receptacle. Ensure
the optical surface is clean before connecting the cable to the lightsource.
®
LIT-217 CUDA
SURGICAL
(English)
This symbol indicates type BF equipment.
If it becomes
Rev. B
Date of Revision: 05/02/17
Page 4 of 71

Publicité

Table des Matières
loading

Produits Connexes pour Cuda surgical Elite 1000

Table des Matières