Apex 9S-006500 Instructions D'utilisation page 16

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

XT Heated Humidifier/ English
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to
earth
Voltage dips,
Voltage Dips:
short
i) 100% reduction for 0.5 period,
interruptions and
ii) 100% reduction for 1 period,
voltage variations
iii) 30% reduction for 25/30 period,
on power supply
Voltage Interruptions:
input lines
100% reduction for 250/300 period
IEC61000-4-11
Power frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V/m 80 MHz to
2.7 GHz
80 % AM at 1 kHz
Radiated RF EM
385-6000 MHz,
Fields
9-28V/m, 80%
IEC61000-4-3
AM(1kHz) pulse
mode and other
modulation
± 1 kV line(s) to
± 1 kV line(s) to
line(s)
line(s)
100 - 240 V
30 A/m
30 A/m
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM and
amateur radio
6Vrms
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
10 V/m 80 MHz to
2,7 GHz
80 % AM at 1 kHz
385-6000 MHz,
10V/m
9-28V/m, 80%
AM(1kHz) pulse
mode and other
modulation
Mains power quality should be
that of atypical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of this device requires
continued operation during
power mains interruptions, it
is recommended that the
device be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications equipment
should be used no closer to
any part of this device,
including cables, than there
commended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d
=
150kHz to 80MHz
d
=0.6
80MHz to 800MHz
d
=1.2
800 MHz to 2.7GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey ,
a
should be less than
the compliance level in each
frequency ranged.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
13
b

Publicité

Table des Matières
loading

Table des Matières