Resistance: This function is only available when test leads are
connected to de-energized target line. Refer to the Direct Test
Leads Connection Mode section 3.5 to properly connect the test
leads. The indicated value is the resistance of the line connected
to the Transmitter. The maximum measured value is 999 kΩ. The >
symbol indicates that the measured value is greater than 999 kΩ.
When at A-Lo / A-Hi mode, the
of a voltage presence ≥10 V (typical) on the circuit under test,
the Ω measurement will be opt out under MENU screen.
Speaker Volume: Use "
then press "ENTER". Use "
the volume. Press "ENTER" to exit the speaker menu.
Contrast: Use "
/
" buttons to highlight the contrast
icon then press "ENTER". Use "
decrease the contrast. Press "ENTER" to exit the contrast menu.

2.5 SC-600 Signal Clamp (included with UAT-620, optional for UAT-610)

In many situations, it is either not possible to gain access to a
cable for making an electrical contact or it is not safe to do so.
The Signal Clamp accessory provides an efficient and safe method
of applying a locate signal to a cable, enabling the Transmitter to
induce a signal through the insulation into the wires or pipes. The
clamp works on low impedance closed circuits only.
indicator will blink. In case
/
" buttons to highlight the speaker
/
" buttons to increase/decrease
/
" buttons to increase/
mA
Ω
10
999 kΩ
ENTER
ENTER
50
ENTER