Schick Dental C2 Master Mode D'emploi page 13

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

5. Setting up
Operating the different versions
Cruising system and operating with digital indicator
see discription on page 6
all versions
5.1 Ascertain, that your mains supply coincides with
the data on the rating system.
5.2 Plug the mains cable 9 into the unit 1 and into
a mains socket.
5.3 Plug the motor cable 3 into the socket 2 on the
rear of the unit (quarter-turn catch).
5.4 Place the handpiece on the rest 8 supplied with it.
knee controller version only
5.5 Press the mains switch 4 on the front of the unit.
The control lamp will come on. The unit is switched.
off by pressing the mains switch 4 again.
5.6 Set the speed controller 5 to the desired speed.
The set speed can be reached right away by
pressing the knee controller 6.
The speed controller is provided with a mechanical
locking which only allows speeds up to 40.000 rpm
unless the segment 5 is operated deliberately.
To reach speeds > 40.000 rpm press segment 5
whilst turning speed controller.
foot controller version only.
5.5 Press the mains switch 11 on the front of the unit.
The LED will flash up. The unit is switched
off by pressing the mains switch 11 again.
5.6 The speed can be varied progressively between
1.000 - 50.000 rpm by moving the foot pedal 13.
For your safety the unit is provided with a mechanical
speed limitation 12. If the red warning light lights up,
up to 50.000 rpm have been enabled. If the red
warning light at the limiting switch is not lit up, the
RPM's are limited to a maximum of 30.000.
bench controller version only
5.5 Plug the dynamic foot-switch 21 into the
socket 23 on the rear of the unit
(quarter-turn-catch).
5.6 Switch on the mains switch 20 on the front of the
unit (the control lamp lights up). The unit is
switched off by pressing the mains switch 20 again.
5.7 Set the speed control knob 22 to the desired speed.
(see digital display 24) The motor can be switched
on or off by moving the pedal and the speed can
be varied progressively up to the set speed.
5.8 The bench unit can also be operated without the dynamic
foot-switch. Set the speed control knob 22 to the
desired speed. The motor will begin to run immediately.