3A HEALTH CARE MAXIASPEED Manuel D'utilisation Et D'entretien page 16

Masquer les pouces Voir aussi pour MAXIASPEED:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

MAXIASPEED does not require maintenance and/or lubrication. Nevertheless, several simple several
simple checks must be performed before every use:
- Check the integrity of the casing, power supply cable and the pedal switch, if supplied.
- Block the aspiration connector with a finger and check that the vacuum level reaches -0.9 bar.
- Check that there are no disturbing noises suspicious of malfunctioning.
- Check that the cage is positioned correctly inside its housing. It must be in axis with the plug aspiration
hole so that the floating valve can block it in the circumstance in which the aspirated liquid exceeds
the maximum volume the container can receive. Do not connect the cage as shown in figure 3a; it must
be connected as shown in figure 3b.
CAGE OUT OF AXIS
- Check that the float is fitted in the correct position and that it is free to move inside the cage (dirt or
encrustation may hinder its movement). Figure 4 shows the correct insertion.
- Check that the plug seal is inserted with the edging as illustrated in figure 5.
FLOAT
Electromagnetic compatibility – additional motes in conformity with the IEC 60601-1-2/A1:2004-09 Standard.
Emission test
RF Emissions
Cispr 11
RF Emissions
Cispr 14
Harmonic Emissions
IEC 61000-3-2
Emissioni di fluttuazioni
di tensione/flicker
IEC 61000-3-3
PERIODICAL CONTROL FOR APPLIANCE SAFETY
Fig. 3a
CAGE
Fig. 4
Conformità
Group 1
--
Classe A
Conforms
Conforme
Emission
Electromagnetic Environment - guidance
MAXIASPEED uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The suction equipment is suitable for use in all establishments
and those directly connected to the public low-voltage power
supply network that supplies buildings used to domestic
purposes. It is possible to use the device in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes, adopting
provisions in the installation, such as a longer distance from
potentially sensitive appliances.
It is possible to use the device in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings user for domestic
purposes.
CAGE IN AXIS
Fig. 3b
Fig. 5

Publicité

Table des Matières
loading

Table des Matières