5 Cleaning, disinfecting and sterilizing
The instructions relating to the accessory cleaning, disinfection and sterilization protocols provided by
Ultradent Products Inc. have been approved for each medical device and accessory. The applicable guides are listed
in chapter Associated documentation page 4
They can be downloaded at the following addresses www.satelec.com/documents or www.ultradent.com
In all cases, the local regulations in force relating to the accessory cleaning, disinfection and sterilization
protocols take precedence over the information provided by Ultradent Products Inc.
5.1 Cleaning and disinfection of the medical device
The medical device must be in OFF or O stop position during cleaning and disinfecting procedures.
Refer to the instructions detailed in the chapter Cleaning the irrigation system page 8
Avoid using cleaning and disinfection products that contain flammable agents.
Otherwise, ensure that the product has completely evaporated from or that there is nothing flammable left on the
medical device and its accessories before switching it on.
Do not use abrasive product to clean the medical device.
Never apply sprays directly to the medical device to clean it. Always spray the product onto a wipe,
then clean the medical device.
The medical device control unit, handpiece cord and control pedal must be cleaned and disinfected daily.
Use alcohol-impregnated wipes for disinfection.
5.2 Cleaning and disinfecting accessories
Refer to the accessory cleaning, disinfection and sterilization protocols listed in the chapter Associated
documentation page 4.
6 Monitoring and maintenance of the
medical device
Before and after each use, check that the device and its accessories are not faulty in any way. This is
necessary to detect any insulation fault or damage. If necessary, replace damaged parts.
Monitor the cleanliness of the air inlets on the control unit to prevent any heating.
7 Maintenance
The only preventive maintenance the medical device requires is:
• checking of accessories;
• everyday cleaning, disinfection and sterilization procedures;
• cleaning.
• replacement of the water filter cartridge.
7.1 Replacing the water filter
The water filter must be replaced or at least cleaned every 6 months.
Proceed as follows:
• shut off the water supply;
• stop the medical device (position O);
• disconnect the network plug;
• unscrew the two filter sections;
J58349 • V1 • (14) • 03/2014 • NB77US010B
9