3 LCSU 4 Setup
Prepare the LCSU 4 for Operation
Note
All models are shipped with the Battery inside the unit, but not connected. Connect the Battery and charge it fully before
using the unit. See Battery Charging (Chapter 5).
Assemble 800 ml Version
A
B
Canister Ports
A - Vacuum Port
B - Patient Port
4
4 Attach the Vacuum Inlet
Connector to the Vacuum
Inlet on the Suction Unit.
Note
The 800 ml Canister has an internal Filter in the Lid. The Canister is disposable and cannot be cleaned. The Filter
automatically stops suction/flow when the Canister is full, or if the LSCU 4 tips over.
Caution
Always use the 800 ml Canister supplied by Laerdal, which has an internal Filter. Never connect any type of Patient Tubing
directly to the LCSU 4 Vacuum Inlet Port Connection. Overflow of suctioned material into the LCSU 4 pump will result in
loss of suction and permanent damage to the unit. In the event of overflow, do not use the LCSU 4. Contact Laerdal Medical
or your authorized distributor.
Operation with High Efficiency Filter Kit
To increase filtration efficiency the Vacuum Tube may be replaced with a High Efficiency Filter Kit
(Cat. No. 886116).
Ensure Filter IN points towards the WHITE connector, and OUT towards the BLUE. For Cleaning
and Maintenance, see Chapter 6.
6
1
1
1 Secure the Lid to the
Canister.
5
5
5 Attach the WHITE
Connector to the Vacuum
Port Connector on the
Canister.
2
2 Place the Canister into
Wire Stand.
6
6 Check that all Vacuum
Tube connections are
firmly attached.
3
3 Ensure that the Patient
Port is accessible.
7
7
7 Attach the Patient Tube
to Patient Port on the
Canister.
IN
OUT