300 ml Disposable Canister
Patient Port
Patient Tube
Wire Stands
Electromagnetic Emissions Tests
Emissions Test
Conducted
and radiated
RF emissions
Harmonic distortion
Voltage
fluctuations/
flicker emissions
Electromagnetic Immunity Tests
Immunity Test
Electrostatic
discharge
Radiated RF EM fields
Proximity fields from RF Wireless
communications equipment
Electrical fast transients / bursts*
Surges Line-to-line*
Conducted
disturbances
induced by RF fields*
Voltage dips*
Voltage
interruptions*
Rated power
frequency magnetic fields
Electrical transient conduction along
supply lines, DC
power port
*AC/DC adapter only.
PC
PP
PVC
Steel, PVC
Standard or test
CISPR 11
IEC 61000-3-2
IEC 61000-3-3
Standard or test
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
IEC 61000-4-8
ISO 7637-2
9 Specifications
Internal Filter: PE
Compliance
Group 1 Class B
Complies
Complies
Compliance Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
10 V/m
80 MHz- 2.7 GHz
80% AM at 1 kHz
380-390 MHz: 27V/m
430-470 MHz: 28V/m
704-787 MHz: 9V/m
800-960 MHz: 28V/m
1700-1990 MHz: 28V/m
2400-2470 MHz: 28V/m
5100-5800 MHz: 9V/m
± 2 kV
± 0,5 kV, ± 1 kV
3 V 0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between
0,15 MHz and 80 MHz
0% UT; 0,5 cycle At 0°, 45°, 90°, 135°, 180°,
225°, 270 ° and 315°
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0°
0% UT; 250/300 cycle
30 A/m
50 Hz or 60 Hz
Test pulse severity level:
III in Table A2 of
ISO 7637-2
15