WAGO Multi- Tester 206-810 Notice D'utilisation page 10

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 14
Volts DC:
RANGE
RESOLUTION
200 mV
0.1 mV
2 V
0.001 V
20 V
0.01 V
200 V
0.1 V
600 V
1 V
Input impedance: 10 MΩ
Max. input voltage 600 V DC
Volts AC
RANGE
RESOLUTION
200 mV
0,1 mV
2 V
0.001 V
20 V
0.01 mV
200 V
0.1 V
600 V
1 V
Input impedance: 10 MΩ
Max. input voltage 600 V AC RMS, frequency range: 40 – 400Hz
Direct current:
RANGE
RESOLUTION
200 μA
0.1 μA
2000 μA
1 μA
20,00 mA
0,01 mA
200,00 mA
0,1 mA
2,000A
10 mA
10,00 A
0,1 A
Overload protection: F 200 mA/690V fuse protection for μA and mA ranges
10 A range is protected by 10 A/690V
Alternating current:
RANGE
RESOLUTION
200 μA
0.1 μA
2000 μA
1 μA
20,00 mA
0.01 mA
200,00 mA
0.1 mA
2,000A
10 mA
10,00 A
0,1 A
Overload protection: F 200 mA/690 V fuse protection for μA and mA ranges
10 A range is protected by 10 A/690 V
Resistance:
RANGE
RESOLUTION
200 Ω
0,1 Ω over-range protection: 250 V DC/AC
2 kΩ
0.001 kΩ
20 kΩ
0.01 kΩ
200 kΩ
0.1 kΩ
2 MΩ
0.001 MΩ
20 MΩ
0.01 MΩ
Measurement voltage: 0.25 V, over-range protection: 600 V AC/DC < 30s
ACCURACY
± 0.5 % of rdg. +3 digits
± 0.8 % of rdg.
+ 5 digits
ACCURACY
± 1.5 % of rdg. + 5 digits
ACCURACY
± 1.0 % of rdg. + 3 digits
± 1.2 % of rdg.+ 5 digits
ACCURACY
± 1.3 % of reading + 5 digits
± 1.5 % of reading + 8 digits
- 10 -
ACCURACY
± 1% + 5 digits
± 1% + 5 digits
± 1.8% + 5 digits