Blue light reading
Blue light PEAK value
5. USING AccuPRO

5.1 ON/OFF

Press and hold the MENU button to turn the meter ON.
Press and hold the MENU button to turn the meter
The Boot Screen (below) will load, followed briefly by
on. Boot screen below followed by Information screen
the Information Screen. (See Section 7 for details about
(section 7) will display. Fully charge before use.
the Information Screen).

5.2 OPERATION MODE

When turned on, the AccuPRO automatically enters
Operation Mode and begins displaying light readings
for white light (555nm) and UV (365nm) wavelengths.
If using the AccuPRO Plus (XP-4000), press the UP
arrow to toggle between this display and blue light
(450nm) readout.

5.3 SENSOR READOUTS AND PEAK VALUE

Moving the sensor head will cause the readout values
to increase or decrease depending on the levels of
illumination, with the highest recorded value stored
automatically as the PEAK value in the lower left of the
readout screen. Images on pages 5 and 6 detail each
readout screen.
Shine the light source being measured directly over
the center of the sensor for the most accurate
readings.
The PEAK value stores the maximum luminosity
recorded since the last reset. Simply press the DOWN
arrow in operational mode to reset the PEAK value to
zero across visible, UV, and if applicable, blue light
readouts.
Operation Mode for Blue Light Sources
6
Battery indicator
Units
5.4 SENSOR WAVELENGTHS & RANGES
The single sensor attached to the AccuPRO will
display UV-A (365nm) readings as well as visible light
(555nm). In addition, the AccuPRO Plus will also
display blue light at the wavelength of 450nm.
Meter will display "XXXX" should readings in any
measured portion of the spectrum exceed the figures
below:
Model No.
Wavelengths Recorded
XP-2000
UV-A/VIS
UV-A (365nm) Irradiance
Visible (555nm) Illuminance
XP-4000
UV-A/VIS/Blue
UV-A (365nm) Irradiance
Visible (555nm) Illuminance
Blue (450nm)

6. CUSTOMIZING SETTINGS

Press MENU, then press the DOWN arrow. Select the
SETTINGS option. Press MENU again to enter the
SETTINGS menu as shown below:
The SETTINGS menu contains the options displayed
below:
Range
0 _ 100 mW/cm
2
0 _ 5,300 Lux
0 _ 500 fc
0 _ 100 mW/cm
2
0 _ 5,300 Lux
0 _ 500 fc
0 _ 100 mW/cm
2

Publicité