Description of your controller's symbols:
There are a number of technical markings on your controller. These can
be explained as follows:
Inside the battery compartment '+' indicates positive polarity and '-'
indicates negative polarity. The batteries should be positioned as indicated
by three embossed outlines in the base of the compartment.
The controller and belt are made in China for Bio-Medical Research Ltd.,
Parkmore Business Park West, Galway, Ireland.
The controller requires three 1.5 volt DC batteries. DC is indicated by the
symbol:
Program
Parameters
14
• Output Current
• Pulse Rate
• Pulse Duration (Width)
• Mode # 1
• Mode # 2
• Mode # 3
• Mode # 4
• Mode # 5
• Mode # 6
• Mode # 7
• Timer
• Compliance
• Waveform
• Maximum Charge Per Pulse
• Power Source
• Tolerances
This symbol means "Attention, consult the accompanying
!
documents".
This symbol means type BF equipment (GelPads).
SN stands for 'serial number'. On the label, on the back of the controller is
the serial number specific to this product. The letter preceding the number
indicates the year of manufacture, where 'U' denotes 2015, 'V' denotes
2016 etc.
The batch number of the belt is represented on the belt packaging,
by the number corresponding with the LOT symbol.
0-70 milliamps, adjustable (up to 1K ohms load)
45, 50, 60, 70 and 75 pulses per second (PPS)
Positive pulse width 200 microseconds / Negative pulse width 200 microseconds /
Inter-phase interval 140 microseconds
Pulse rate 70 PPS / On time = 2.5 seconds / Off time = 2.5 seconds
Ramp Up and Down = 2 seconds / Timer = 20 minutes
Pulse rate 60 PPS / On time = 3.5 seconds / Off time = 3.5 seconds
Ramp Up and Down = 2 seconds / Timer = 25 minutes
Pulse rate 50 PPS / On time = 4.5 seconds / Off time = 4.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
Pulse rate 50 PPS / On time = 5.5 seconds / Off time = 5.5 seconds
Ramp Up and Down = 2 seconds / Timer = 40 minutes
Pulse rate 75 PPS / On time = 6.5 seconds / Off time = 6.5 seconds
Ramp Up and Down = 2 seconds / Timer = 30 minutes
Pulse rate 70 PPS / On time = 10 seconds / Off time = 10 seconds
Ramp Up and Down = 4 seconds / Timer = 30 minutes
Pulse rate 45 PPS / On time = 10 seconds / Off time = 10 seconds
Ramp Up and Down = 4 seconds / Timer = 30 minutes
20, 25, 30 and 40 minutes depending on the mode.
Complies with IEC 60601-1
Symmetrical biphasic with zero net DC
15 microcoulombs under normal conditions
3 x 1.5V AAA Batteries
All electrical specifications are ± 10% into real 500-ohm load
English
www.slendertone.com