DeVilbiss Healthcare SmartLink DV5M Série Manuel D'utilisation page 12

Masquer les pouces Voir aussi pour SmartLink DV5M Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
Immunity Test
Power Frequency
50/60Hz Magnetic
Field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
For transmitters rated at a maximum output power not listed above, the recommended separation distance
D in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
EN - 12
IEC 60601 Test
Level
3A/m
3 Vrms from
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance Level
3A/m
V1 = 3 Vrms
E1 = 3V/m
Electromagnetic Environment -
Guidance
Power frequency magnetic fields
should be that of a typical location in
a typical commercial or hospital
environment.
Portable and mobile RF
communications equipment should be
separated from the device by no less
than the recommended separation
distances calculated/listed below:
D=(3.5/V1)√ P
D=(3.5/E1)√ P
80 to 800 MHz
D=(7/E1)√ P
800 MHz to 2.5 GHz
Where P is the maximum power
rating in watts and D is the
recommended separation distance in
meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey, should
be less than the compliance levels
(V1 and E1).
Interference may occur in the vicinity
of equipment containing a transmitter.
A-DV5M-1

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Smartlink dv5m

Table des Matières