Immunity test
IEC 60601 test
level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
1.
U
is the ac mains voltage prior to application of the test level.
T
2.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the
EUT is used exceeds the applicable RF compliance level above, the EUT should be observed
to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorien-
ting or relocating the EUT.
3.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Note 1: At 80Hz and 800MHz, the higher frequency range applies.
Note 2: These guidances may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Compliance
Electromagnetic environment –
level
Guidance
3 V/m
Recommended separation
distance:
d = 3,5/3 * SQRT (P)
d = 3,5/3 * SQRT (P) 80 MHz to
800 MHz
d = 7/3 * SQRT (P) 800 MHz to
2,5 GHz
where „P" is the maximum output
power rating of the transmitter in
watts (W) according to the transmit-
ter manufacturer and d is the recom-
mended separation distance in
metres (m).
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey
less than the compliance level in
each frequency range
ence may occur in the vicinity of
equipment marked with the follow-
ing symbol:
Table 2 and 4 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007)
2)
, should be
3)
. Interfer-
183