mecHanicaL inStaLLation
All sensors come with a 1m cable and a mounting nut. The sensor can be mounted on brackets
or cut outs of your own design, or you can use the fixed and adjustable
mounting bracket accessories which are shown below. Note: The PyroCouple M sensor housing
must be connected to earth.
Fixed Bracket
45.0
50.0
2 x Mounting Holes M4 Clearance
24.0
9.0
air purge coLLar
The air purge collar below is used to keep dust, fumes, moisture, and other contaminants away
from the lens. It must be screwed in fully. Air flows into the 1/8" BSP fitting and out of the front
aperture. Air flow should be no more than 0.5 to 1.5 litres/min.
Clean or 'instrument' air is recommended.
APSW
Ø16.0
12.0
25.0
60° Rotation
1/8 BSP Air connection
25
50
APSN
Adjustable Bracket
9.0
40.0
50.0
2 x Mounting Holes M4 Clearance
24.0
9.0
60° Rotation
15.0
60° Rotation
25.0
48.0
20
29
40
5