DURING/AFTER POWER FAILURE
Display of chamber temperature during power failure
By pressing the buzzer stop key (BUZZER) during a "power-failure alarm", the buzzer stops and the
chamber temperature is displayed on the temperature display for 5 seconds.
Check the chamber temperature as appropriate by pressing the buzzer stop key (BUZZER) during
power failure.
<Important>
The battery for the power-failure alarm has a limited lifespan. Replace the battery every 3 years. The
alarm may not be activated if the power fails and the battery has not been replaced regularly. Contact
our sales representative or agent to arrange replacement of the battery.
Operation check after recovery from power failure
After recovery from a power failure, the operation will resume automatically with the same settings as
before the power failure.
There is no need to re-set.
However, always check the operational
status after recovery.
The settings are memorized during power failure by nonvolatile memory.
<Important>
Upon recovery from a power failure, check that the unit restarts operation properly. Also check that
there are no changes to the settings.
20