Problem
Display works, but no
counts are registered
Reading is high, but
another instrument;
has a normal reading
in the same location
Instrument has false
high reading
Instrument has false
high reading
Instrument has false
high reading
Instrument has false
high reading
Service
CAUTION: Do not send a contaminated instrument for repair or calibration under
any circumstances. There are no user serviceable parts inside
instrument.
If the Inspector+ requires servicing, please contact your distributor or the