4. Using ultrasound mode
Choosing the ultrasound power range
1.
Select a power range by pressing one of the
four keys (green, yellow, blue, orange) (ref.
16) on the keyboard.
By default the display indicates the color of the
range selected at power setting 5 (1 for high
power) and irrigation from the water supply.
Reminder: the ultrasound power range may be selected
with the multifunction footswitch.
Setting ultrasound power
Incremental setting
1.
Press the
key to increase ultrasound power
or the
key to decrease it from 1 to 10, by
steps of 1.
Setting the power to maximum (
1.
Press the
key on the front of the unit (ref.
17) or the multifunction footswitch (ref. 40) to
set the power to maximum.
The indicator lights up (ref. 9) and the display
indicates the value of 10 in the selected range.
To exit Boost mode, press the key on the front of
the unit or press the multifunction footswitch
once more, or change the power range.
Spray or dry
1.
Press the
key on the face of the unit. (ref.
21) or the multifunction footswitch (ref. 41) to
activate irrigation.
In water supply mode, controlling the water
-
flow is via the connector ring on the
®
Suprasson
handpiece (ref. 55)
In irrigation kit mode, controlling the water
-
flow is via the flow limiter located on the
pump (ref. 46).
Memorizing the ultrasound power
level
1.
Select the desired power range.
2.
Change the power level with the
(ref. 22)
3.
Press the memory key
The power display shows "--" for one second to
confirm memorization.
Recalling a level from memory
1.
Press the key for the desired power range (ref.
16)
The memorized value is displayed.
Refer to the Tipbook for power settings of
Satelec tips.
)
keys
. (ref. 18)
7