Tds Parameter; Instrument Setup Menu - Dostmann pH 7 Vio Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

TDS Parameter

COND 7 Vio, PC 7 Vio
The Conductivity measurement can be converted into the TDS Parameter.
This parameter uses the Conductivity calibration; therefore, refer to the previous paragraph to
calibrate the sensor.
Total Dissolved Solids (TDS) correspond to the total weight of the solids (cations, anions and non-
dissociated substances) in a liter of water. Traditionally, TDS are determined using the gravimetric method,
but a simpler and faster method is to measure Conductivity and convert it to TDS by multiplying it by the
TDS conversion Factor.
In measure mode press
Use the directional keys to move to TDS SETTINGS P4.0 and access the menu by pressing the key
.
Press again
to access the TDS FACTOR P4.1 program.
When the value flashes, use the directional keys as indicated by the icon
value and confirm with
By default, the TDS factor is set at 0.71; the user can change it between 0.40 ... 1.00.
Here below, the TDS factors in relation to the Conductivity value are shown:
The TDS measurement is expressed in mg/l or g/l depending on the value.

Instrument Setup menu

• In measure mode, press key
• Use the directional keys to move to SETTINGS P9.0 and access the menu by pressing the key
• Move with the keys
The table below shows the setup menu structure for the general settings of the instrument; for each
program, there are the options that the user can choose and the default value:
Program
Description
P9.1
TEMPERATURE U.M.
P9.3
BACKLIGHT MODE
P9.4
BRIGHTNESS
P9.5
SLEEP MODE
P9.6*
SELECT PARAMETER
P9.8
RESET
P9.9
AUTO POWER-OFF
* Function available for PC 7 Vio only.
Series 7 Vio ENG Version 2.0 April 2020
to access the SETUP menu.
.
Conductivity of the Solution
1-100 µS/cm
100 – 1000 µS/cm
1 – 10 mS/cm
10 – 200 mS/cm
to access the SETUP menu.
and
to select the program to access.
°C / °F
INDOOR – OUTDOOR
LOW – NORMAL - HIGH
OFF – 2 MIN – 5 MIN
YES – NO for each parameter
YES - NO
YES – NO
30
to enter the correct
TDS Factor
0.60
0.71
0.81
0.94
Options
.
Factory Default
Settings
°C
INDOOR
NORMAL
2 MIN
YES
NO
NO
EN