Temperature measurement ATC – MTC
ATC : The direct measurement of the sample temperature for all parameters is carried out through
the NTC 30KΩ probe, which can be either integrated into the sensor (electrode and / or cell) or
external.
MTC : If no temperature probe is connected, the value must be changed manually:
keep pressed
or
directional keys; then press

pH Parameter

pH 7 Vio; PC 7 Vio
On this series of devices, it is possible to use pH sensors with integrated temperature probe or to connect
two different sensors. Connect the pH electrode to the BNC type connector marked in green. Connect the
temperature probe to the RCA / CINCH Temp connector always marked with a green background.
The instrument is also able to recognize the DHS sensor, an innovative electrode that stores calibration
data and that can be used immediately after on any enabled instrument.

• pH parameter Setup

In measure mode press
Press the button
Move with the keys
The table below shows the setup menu structure for the pH parameter, and for each program the options
that the user can choose and the default value:
Program
Description
P1.1
CAL BUFFER SELECT
P1.2
SELECT RESOLUTION
P1.3
STABILITY CRITERIA
P1.6
CALIBRATION DATA
P1.8
RESET SETTINGS
P1.9
TEMPERATURE CAL
P1.1 pH buffer selection
Access this setup to select the buffer family with which to perform the electrode calibration.
Calibration lines from 1 to 3 points can be made.
During calibration, press
(see calibration paragraph).
The instrument automatically recognizes 2 families of buffers (USA and NIST); in addition, the user has the
option of performing a manual calibration of up to 2 points with customizable values.
Series 7 Vio ENG Version 2.0 April 2020
until the value starts to flash; then adjust it by continuing to use the
to confirm.
to access the SETUP menu.
to access the pH SETTINGS P1.0 menu
and
to select the program you want to access.
USA – NIST – Custom
0.1 – 0.01
LOW – MEDIUM - HIGH
-
YES – NO
YES – NO
to exit and save the points calibrated up to that moment
17