Accu-Chek Tender Instructions D'utilisation page 7

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

• Dispose of your infusion set after use according to local regulations. Do not
clean or re-sterilize.
• Reuse of the infusion set may cause infection, site irritation or damage to the
cannula/needle. A damaged cannula/needle may lead to inaccurate insulin
delivery.
• The infusion set is for subcutaneous use only.
• Always screw the luer-lock connector firmly into the pump adapter. If this is
not done, a loose connection between the infusion set and the pump could
cause leakage. Avoid using tools to screw the luer-lock connector into place,
otherwise damage could result to the luer-lock connector.
• Do not leave air bubbles in the infusion set. Prime thoroughly.
• A small percentage of the Accu-Chek Tender infusion sets using soft cannulas
can accidentally become crimped during insertion or displaced while in use.
If this should occur, use a new infusion set immediately.
• Check the infusion site frequently to make sure the soft cannula remains
firmly in place. Because the cannula is so soft, you will not notice any pain if
en
6

Publicité

Table des Matières
loading

Table des Matières