Inspection Schedule
Weekly inspection:
DePuy Mitek recommends a weekly inspection of
all equipment. These inspections can be performed
on a less frequent basis depending on the duration
and frequency of use.
- P erform a visual inspection of the device. If
damage is noted, contact your DePuy Mitek
Sales Representative or call DePuy Mitek
Customer Service in the USA at +1-800-382-
4682. Select Option 2 to speak to a biomed.
- I nspect that the transparent safety cover (1) and
black hinges are in good working order.
- I nspect power cord and all cables for wear or
cuts.
- E nsure that the foot pedal and shaver interface
connector pins are not missing or bent.
- P lug the power cord to the power connection.
- S witch the I/O switch (14) on "I" .
- T urn the device ON by pressing the POWER (4)
pad.
Press on all pads. An audible tone should be heard
for each pad pressed, except for Pressure Cycle,
(Pad 8) Pressure Cycle is inactive until the pump is
primed.
109076-KIM Rev B.indd 8
Annual inspection:
DePuy Mitek recommends that a bio-medical
engineer or authorized service personnel annually
check all equipment to assess its functionality and
technical safety.
8
9/26/12 4:02 PM