3.2. Tubing Set-up.

WARNING: This product is for single-use only. It
has not been designed to be re-used/re-sterilized.
Reprocessing may lead to changes in material
characteristics such as deformation and material
degradation which may impact the strength of the device
and compromise device performance. Reprocessing of
single use devices can also cause cross-contamination
leading to patient infection. These risks may potentially
affect patient safety.
Sterility is preserved only if the tube changing
procedures have been performed precisely according to
the set-up instructions below and in Section 3.2.E.
This product contains DEHP, a phthalate chemical
known to cause birth defects and other reproductive
(A)
Clamps
From
solution
bags
Clamps
(A)
Close all three clamps and
connect to fluid bags.
109076-KIM Rev B.indd 17
harm, including in children and in pregnant or nursing
women.
This tube connects the saline solution bags to the
irrigation tubing. Solution goes from the spikes (A)
around the roller pump, through the pressure chamber,
where the pressure is read through a filter (B) by a built-
in pressure transducer.
A. Installing the One Day Set Irrigation Tube
Ref. 284504 or 284503
Contact your DePuy Mitek Representative for ordering
information since certain tubes are not offered in all
countries.
GREEN
Auto-locking device
(B)
To the Pressure
Filter
transducer
Slide the green part of the
tube into the auto-locking
device on the pump.
17
(C)
PRESSURE CHAMBER
Clamps
Place and center the tube
around the roller pump.
Intermediary
Irrigation Tube
9/26/12 4:02 PM