wiring details
2
Stecker / Plug / Fiche
1 = braun, 2 = weiß, 3 = blau, 4 = schwarz / 1 = brown, 2 = white, 3 = blue, 4 = black / 1 = brun, 2 = blanc, 3 = bleu, 4 = noir
wiring details
element
Typ J (Fe-CuNi)
Typ T (Cu-CuNi)
Typ K (NiCr-Ni)
Typ E (NiCr-CuNi)
Typ N (NiCrSi-NiSi)
Typ S (Pt10Rh-Pt)
Typ R (Pt13Rh-Pt)
Typ B (Pt30Rh-Pt6Rh)
Between 850 and 950 °C type K involves the risk of selective chrome oxidation. If the service temperature
is continually fluctuating within this area, we recommend sensor type N.
element
Typ L (Fe-CuNi)
Typ U (Cu-CuNi)
TS with Pt100 L-plug M12x1
1
ϑ
4
3
3-Leiter, 3-wire, 3-fils
1 x thermocouple
thermocouples
thermocouples
4
3
ϑ
1
4-Leiter, 4-wire, 4-fils
IEC 584-1
min.
max.
-210°C
760°C
-270°C
400°C
-270°C
1370°C
-270°C
1000°C
-270°C
1300°C
-50°C
1760°C
-50°C
1760°C
+100°C
1820°C
DIN43710
min.
max.
-200°C
900°C
-200°C
600°C
1
2
3
4
2 x thermocouple
+ connection - connection
black
brown
green
violet
white
lilac
orange
orange
grey
+ connection - connection
red
blue
red
brown
Quality for the
future