4. Safety functions
4.1. Emergency stop
On the left hand side, you will find the emergency stop
(See figure 23). Operate this button in an emergency; if
the upward or downward movement must be stopped
immediately. Once the emergency stop has been pressed,
the electronic controls will stop working. Operation may
be continued once the situation is stable again. Turn the
emergency stop a quarter turn to unlock it again (See
figure 24).
Check whether the danger has passed or the problem
has been resolved before resetting the emergency stop.
The orange LED on the hand control will flash when
the emergency stop has been activated.
4.2. Sleep setting
After around an hour, the chair will fall into a 'sleep' setting
and the battery indication LEDs on the manual control will go
out. In order to use the chair again, it must be activated (See paragraph 2.2. Activating).
4.3. Emergency lowering
If the controls stop working and the chair is in a raised position,
it can be lowered using the emergency lowering button.
Check first to make sure the chair is not in sleep setting!
• Remove the plug from the manual control (1) (See figure 25).
• Press the emergency lowering button (2) (See figure 25).
The chair will now return to the lowest position (make sure
no persons or objects are under the chair). Take the client
from the chair and make sure the device is not used until it
has been repaired. Place the chair in a safe place and indicate that it is out of order.
Contact the technical service or Lopital's service department.
E-mail: service@lopital.nl
38 | Instructions for use | Elexo Shower-toilet chair
Figure 23 Emergency stop.
Figure 24 Emergency stop.
2
1
Figure 25 Emergency lowering
button.