Methods
GB
Notes:
1. This test is applicable for the determination of soluble Manganese in water and waste-
water.
2. Highly buffered water samples or extreme pH values may exceed the buffering capacity
of the reagents and requires sample pre-treatment.
If samples were acidified for storing, adjust the pH between 4 and 5 with 5 mol/l (5 N)
Sodium hydroxide before test. Do not exceed pH 5, as manganese may precipitate.
3. Interferences:
Interfering substance
Calcium
Chloride
Iron
Magnesium
4. Conversion:
mg/l MnO
= mg/l Mn · 2.17
4
mg/l KMnO
= mg/l Mn · 2.88
4
Reagent
Set
VARIO Manganese Citrate Puffer F10
VARIO Sodiumperiodate F10
AL100_7a 05/2019
Interference level
greater than 700 mg/l
greater than 70 000 mg/l
greater than 5 mg/l
greater than 100 000 mg/l
Form of reagent/Quantity
Powder Pack / 100
Powder Pack / 100
Order-No.
4535100