BCM
8 Menu Sequence Overview
Main menu
Submenu 1
Configuration
H__i
Shortcut
+
Equipment configuration
or 60s no action:
Exit main/submenu
____
Display
____
Measurement display
Humidity
HuMi
____
Out 1
out1
____
Humidity main menu
Settings menu OUT1
____
Out x
out_
Settings menu OUTx
____
Analog Out1
An1
Settings menu Analog 1
Exit
E
Exit menu level
Temperature
tEMP
Out X
out_
Temperature Menu
Settings menu OUT X
Out X
out_
Settings menu OUT X
Analog Out2
An2
Settings menu Analog 2
Exit
E
Exit menu level
Basic EF
b. E F
Reset
rES
Basic Adv. Functions
Factory setting
Errh
Set error handling
Error handling
Set Unit Humidity
Huni
Humidity Unit
t. u ni
Set Unit Temp
Temperature unit
Set Upper
100
Set 100 %
Set Lower
0
Set 0 %
reassign O1
r. o u 1
Set Out 1
reassign O2
r. o u 2
Set Out 2
r. o u_
reassign Ox
Set Out x
Display Rate
di5
Display rate
Loc
Lock Device
Keylock
Exit
E
Exit menu level
Diagnostic
diA
Journal Out
J. o ut
Diagnostics menu
Log output
Journal Error
J. E rr
Logs errors
Max/Min Humidity
H. M M
Max/Min memory
Max/Min Temp
t. M M
Max/Min memory
Set Journal Out
SJ. o u
Output monitoring
Delay MM
DH. M M
Min/Max delay
Delay MM
dt. M M
Min/Max delay
E
Exit
Exit menu level
Exit
E
Exit menu level
BX150101 ◦ 09/2020
Submenu 2
Submenu 3
SP1
____
FH1
F1. H i
____
Switching point 1
rP1
____
FL1
F1. L o
____
Switch-back point 1
____
____
EF 1
EF1
Delay SP 1
dS1
Extended functions
Delay OUT1 On
Delay RP 1
dr1
____
Delay OUT1 Off
____
Set Out 1
ou1
Definition OUT1
____
Check Out 1
C. o u1
Functional test OUT1
____
LED Out 1
LEd1
Function status LED1
E
____
Exit
Exit menu level
____
Exit
E
Exit menu level
See menu structure Out1
Not applicable for version with
analogue output
____
Analog Max1
A1. H i
Set maximum value
____
Analog Min1
A1. L o
Set minimum value
____
Set A.Out1
A. o u1
Output function
____
Check An1
C. A n1
Functional test Analog 1
E
____
Exit
Exit menu level
See menu structure Out1
See menu structure Out1
Not applicable for version with
analogue output
See menu structure Analog Out1
Settings and Display Menu
____
Switch
____
Switches the measurement display level / temperature
Switch view
____
Selection menu
____
Switching point setting /
FH1 Define maximum window limit
SP1
< o. L o - o. H i >
FI. H i
Sets a value for frequency 100 Hz (only OUT1 and OUT2)
____
Selection menu
____
rP1
Set switch-back point /
FL1 Set lower window limit
FI. L o
Sets a value for frequency 1 Hz (only OUT1 and OUT2)
< o. L o - o. H i >
____
Selection menu
___
Amount of time the signal must be present for output OUT1 to trigger.
0 - 100
Setting range: 0...100 seconds (for frequency output without function)
Selection menu
____
___
Amount of time the signal must be present for output OUT1 to switch back.
0 - 100
Setting range: 0...100 seconds (for frequency output without function)
Definition of switching characteristics for output 1:
____
Selection menu
___
Hno Hnc Hysteresis function NO contact or NC contact
Hno - FM
Fno Fnc Window function NO contact or NC contact /
n. o P
Normal switching output operation
____
Selection menu
____
oFF
Switching output Off /
F1
n. o p - on
on
F100
Switching output On /
____
Selection menu
L=__
LED = output: follows the switching output
L= o
L= o - L=-o
L=-o
LED = -output: inverted to switching output
____
Selection menu
____
Sets the maximum analogue output signal to a defined value.
(Note: Max. – Min >= 10 % of the measuring range)
0; C - 100; C
____
Selection menu
____
Sets the maximum analogue output signal to a defined value.
(Note: Max. – Min >= 10 % of the measuring range)
0; C - 100; C
____
Selection menu
_ _
Selecting the analogue output function:
i I
4...20 mA
u 1
2...10 V
i 1 - u 3
____
Selection menu
____
n. o P
Analogue output normal mode /
HALF
Output mean analogue value /
n. o P - Lo
Alternate menu for version with analogue output
Not applicable when ordered with 2 or more
switching outputs
Alternate menu for version with analogue output
Not used when ordered with 4 switching outputs
____
Confirm
___
Resets the device to delivery status
no - YES
____
Selection menu
Selection menu
___
Disable/enable normal error handling and analysis
no - YES
____
Selection menu
___
Selecting the unit displayed:
-1- [Percent],
-l- - non
____
Selection menu
; _
Selecting the temperature unit displayed:
; C - ; F
____
Selection menu
____
Sets a defined value for the maximum humidity.
(Example: maximum humidity at 100 %)
-999 - 9999
____
Selection menu
____
Sets a defined value for the minimum humidity.
(Example: minimum humidity at 10 %)
-999 - 9999
____
Selection menu
____
Set switching output Out1 to humidity measurement/temperature measurement or as
HuMi- tEMP - Err
error signal.
____
Selection menu
____
Set switching output Out2 to humidity measurement/temperature measurement or as
HuMi- tEMP - Err
error signal.
Selection menu
____
Sets switching output OutX to humidity measurement or temperature measurement
HuMi- tEMP
____
Selection menu
____
Selecting the display refresh rate:
FASt - off
FASt [fast], Medi [medium], SLo [slow],
____
Selection menu
___
Activating the keylock by entering a numerical code > 0 Master code 287
0 - 999
____
Selection menu
____
Displays the last 6 alarms + delete function.
Jor I - ---
Display:
JorX
X – amount of time passed in hours/days; scroll through journal with ▲▼
____
Selection menu
____
Displays the error codes + delete function.
E___
Display: Error code
; scroll through journal with ▲▼
____
Selection menu
____
Displays the limits + delete function.
___; _ - ---
Display: Value – amount of time passed in hours/days; scroll through journal with ▲▼
____
Selection menu
____
Displays the limits + delete function.
Display: Value – amount of time passed in hours/days; scroll through journal with ▲▼
___; _ - ---
____
Selection menu
out_
Selecting the alarm output to be logged
Out1 - Out_
____
Selection menu
___
Amount of time the signal must be present before saving the maximum / minimum
value. Setting range: 0...100 seconds
0 - 100
____
Selection menu
___
Amount of time the signal must be present before saving the maximum / minimum
value. Setting range: 0...100 seconds
0 - 100
FM
Frequency output 1-100 Hz
Simulation 1 Hz
Simulation 100 Hz
u 2 0...10 V
u 3 0...5 V
Hi Output maximum analogue value
Lo Minimum analogue value output
non [no unit (4-digit display)]
; C [Celsius],
; F [Celsius]
oFF [Display off]