Télécharger Imprimer la page

Philips Respironics 1126942 Mode D'emploi page 13

Publicité

Guidance and Manufacturer's Declaration - Electromagnetic Immunity – This battery pack
is intended for use in the electromagnetic environment specified below. The user of this
battery pack should make sure it is used in such an environment.
Immunity
IEC 60601 Test Level
Test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
6 Vrms
Amateur Radio & ISM
Bands between 150 kHz
and 80 MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
10
Specifications
Battery Pack Service Life
Under normal conditions, the battery pack is expected to last 3 years.
The following factors can potentially reduce the life span of your battery pack:
• Not inspecting the battery pack for damage before using it or after cleaning
• Improper cleaning or storage
• Storing the battery pack in conditions outside of the temperature ranges listed in this
document
Detailed Specifications
Environmental
Operation
conditions
Storage temperature
Physical
Dimensions
Weight
Compliance Level
3 Vrms
150 kHz to 80 MHz
6 Vrms
Amateur Radio & ISM
Bands between 150 kHz
and 80 MHz
10 V/m
Operating
temperature:
Relative
humidity:
Atmospheric pressure range:
Electromagnetic Environment -
Guidance
Portable and mobile RF communications
equipment should be used no closer
to any part of the device, including
cables, than the recommended 30 cm
separation distance.
Interference may occur in the vicinity of
equipment marked with the
following symbol:
5° C to 40° C (41° F to 104° F)
ambient
15% to 95%, non-condensing
101.3 kPa to 76.7 kPa (0–7,500 feet
above sea level)
-20°C to 60°C (-4°F to 140°F)
22 cm x 13 cm x 2.9 cm
(approx. 8 5/8 in. x 5 1/16 in x 1 1/8 in)
1 kg (2.2 lbs)