Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips and
interruptions
IEC 61000-4-11
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle
and
70 % UT; 25/30
cycles
Single phase: at 0°
0 % UT; 250/300
cycle
± 8 kV contact
± 15 kV air
Not applicable
Not applicable
Not applicable
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30 %.
Not applicable
Not applicable
Not applicable