M Air T Calibration Kit User Guide

Introduction

The M Air T™ Calibration Kit allows you to check and calibrate
the Millipore M Air T Tester with an accuracy of +/-5%. Milli-
pore calibrates the calibration kit before delivery with an accu-
racy of +/-2.5%.
NOTE: Millipore recommends that you send the calibration kit
at least once a year to our closest subsidiary to calibrate
it against a reference flowmeter.

Calibration Kit Components

The calibration kit includes:
■ A float type flowmeter
■ A calibration sieve
■ A calibration certificate (including conversion tables plus
the table and flow rate curve equation specific to your kit)
■ A calibration key
NOTE: Optional calculation software can be provided by our
technical department.