Table Des Matières; Intended Use; Warnings - Mangar Airflo 24 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Airflo 24
Contents

Intended Use

The Airflo 24 by Mangar Health is a low pressure
air compressor, which has been designed to
inflate the Mangar 'Camel' and 'ELK' pneumatic
lifting equipment through a 4-Way pneumatic
hand control. The 4-Way hand control is
supplied with the Camel and ELK.
Applied Parts (Accessories) for the Airflo 24:
Mangar ELK (Emergency lifting cushion)
Mangar Camel (Emergency lifting cushion)
The ELK requires assistance to help lift a patient
from the floor to an elevated sitting position.
Once in a sitting position, the patient can be
transferred to a chair or helped to a standing
position.
The Camel can be operated with or without
assistance and when inflated, lifts the patient
into a raised seated position ready to stand.
No special skills or training are required to
operate the Airflo 24 and its accessories.
Further guidance on operation can be found in
the individual accessory User Instructions.
The Airflo compressor and applied parts are
intended for continual re-use as required and
no special actions or operations are needed to
achieve this.
This product and associated packaging does not
contain latex.
6
Pg 6
Pg 6
• Be sure that you have read and understood these
Pg 7
instructions and those provided with any other
Mangar Health product, before operating the
Pg 8
equipment.
Pg 9
Pg 9
• To avoid the risk of electric shock, NEVER recharge
the battery in the bathroom, outdoors or in a
Pg 10
damp environment.
Pg 10
Pg 11
• Always disconnect from the battery charger
before cleaning.
Pg 11
Pg 12
• Use only the supplied battery charger.
Pg 13
Pg 13
• Do not operate the Airflo 24 compressor in the
presence of flammable anaesthetic mixtures with
Pg 50
air, oxygen or nitrous oxide, as these mixtures
can create a combustible environment leading to
the potential for fire or even an explosion in the
presence of a spark.
• Ensure that when in use, the Airflo 24 and its
connected hoses are laid out in such a way as to
avoid any risk of tripping or strangulation.
• Ensure that children and pets do not interfere
with the safe operation of the Airflo 24 and
inflatable product.
• Keep the Airflo 24 and inflatable product away
from sources of heat.
• Disconnect the Airflo 24 compressor from the
inflatable product before transporting so as to
avoid accidental inflation.
• The Airflo 24 should only be used for its stated
purpose and should not be modified in any way.
• The optional mains power adaptor is specific to
the Airflo 24 and is the only approved unit to be
used for this purpose.
• Do not operate the compressor outside of the
specified duty cycle of 20% 'Operating Time' and
80% 'Non-Operating Time'. Maximum continuous
operating time = 5 minutes.
• To calibrate the battery and battery indicator, it
is essential before the first use to charge for 12
hours and then disconnect from the charger
and leave for 90 minutes to allow the battery
management software to correctly calibrate.
Failure to do this could lead to incorrect battery
charging and battery ladder indication.

WARNINGS

Publicité

Table des Matières
loading

Produits Connexes pour Mangar Airflo 24

Table des Matières