Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10 Faults

Table 2: Faults
Problem
The FMX has been
switched on, but when it
approaches an
electrostatically charged
object to be measured, the
reading does not react
The zero button has been
pressed, but the FMX
cannot be reset
The FMX has been
switched on, but the
display shows no symbols
One of the symbols on the
display is lacking, or the
display shows Error ("Err")
Buzzer does not operate
Voltage measured at ion
balance is too low
FMX-004_UM_9752020710_NL_D_GB_F_V1_0
Cause
The FMX is in the HOLD
mode
The FMX is in the HOLD
mode
Internal zero point has
shifted
Battery voltage too low
Display is defective
Display or sensor is
defective
Faulty buzzer
Dirt or moisture has
bonded to the adapter
needed to measure the ion
balance.
Remedy
Switch the HOLD mode off
Switch the HOLD mode off
Return FMX for calibration
Replace the battery
Replacement or repair is
necessary, send the FMX to
Simco-Ion for repair
Replacement or repair is
necessary, send the FMX to
Simco-Ion for repair
Replacement or repair is
necessary, send the FMX to
Simco-Ion for repair
Clean the adapter with
ultrasound and dry thoroughly.
Send the plate to Simco-Ion for
repairs if this does not help.
GB
43

Publicité

Table des Matières
loading

Produits Connexes pour Simco FMX-004

Table des Matières