Clinical Safety; Electromagnetic Compatibility (Emc) - FujiFilm Sonosite SII Stand Guide D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 57
Cautions:

Clinical safety

WARNING:

Electromagnetic compatibility (EMC)

See the safety section of the ultrasound system user guide and supplemental guides for information on
electromagnetic compatibility with IEC 60601-1-2, including the Manufacturer's Declaration.
14
Excessive bending or twisting of cables can cause a failure or intermittent
operation.
Improper cleaning or disinfecting of any part of the accessories and peripherals
can cause permanent damage. For cleaning and disinfecting instructions, see
"Cleaning and disinfecting"
instructions.
Use only peripherals recommended by FUJIFILM SonoSite.
Accessible metal on the signal inputs and outputs on the back and side of the
ultrasound system is not protectively earthed. Do not perform a high current
ground impedance test involving this part.
To avoid misdiagnosis, do not use non-medical (commercial) grade peripheral
monitors. These monitors have not been verified or validated by FUJIFILM
SonoSite as being suitable for diagnosis.
on page 11 and the manufacturer's operating