Annexe B Modbus Register Information - Hach 3400 Manuel D'utilisation

Table des Matières

Publicité

Annexe B
Group Name
Tag Name
Tags
Conductivity
Tags
Temperature
Measurements
Conductivity
Measurements
Temperature
Settings
MeasMin
Settings
MeasMax
Settings
MeasFormat
Settings
MeasUnitsCond
Settings
MeasUnitsResist
Settings
MeasUnitsTDS
Settings
MeasUnitsSalinity
Settings
TempUnits
Settings
Parameter
Settings
DisplayFormat
Settings
Filter
Settings
TDSConfig
Settings
TDS Factor
Settings
Cell Constant
Settings
Cell Constant Min
Settings
Cell Constant Max
Settings
CellConstSel
Settings
TCompSlope
Settings
TCompRefTemp
Settings
TElementType
Settings
TElementFactor
Settings
TElementManual
Settings
OutPutMode
Calibration
Cal Value
Settings
Sensor Name
Diagnostics
Driver Version
Diagnostics
Serial Number
Modbus Register Information
Tableau 13 Sensor Modbus Registers
Register #
40001
40002
40003
40005
40007
40009
40011
40013
40014
40015
40016
40017
40018
40019
40020
40021
40022
40024
40026
40028
40030
40033
40035
40041
40042
40048
40050
40052
40054
40060
40068
Data Type
Length
R/W
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Float
2
R
Float
2
R
Float
2
R
Float
2
R
Unsigned
2
R
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Unsigned
1
R/W
Integer
Float
2
R/W
Float
2
R/W
Float
2
R/W
Float
2
R/W
Unsigned
1
R/W
Integer
Float
2
R/W
Float
2
R/W
Unsigned
1
R/W
Integer
Float
2
R/W
Float
2
R/W
Unsigned
1
R/W
Integer
Float
2
R
String
6
R/W
String
8
R/W
String
6
R/W
Description
Sensor meas tag index
Temperature tag index
Sensor measurement
Temperature measurement
Minimum meas. value
Maximum meas. value
Display format
Siemens units
Ohm units
TDS units
Salinity units
Temperature units
Selected primary parameter
User selected display format
Number of samples to average
TDS configuration
TDS multiplier
Cell constant value
Minimum cell constant value
Maximum cell constant value
Cell constant selection:
0.01, 0.05, 0.1, 0.5, 1.0, 5.0, 10.0
Temp. comp. slope
Temp. comp. ref. temp
Temp. element:
Manual, Pt100, Pt1000 = 0/1/2
Temp. element offset
Temp. manual temperature
Output mode during calibration:
Active/Hold/Transfer = 0/1/2
Calib. value
Name of sensor
Version of driver
Sensor serial number
43