safety information
SPECIFIC WARNINGS
!
TO REDUCE THE RISK OF ELECTRICAL SHOCK OR INJURY TO PERSONS, DO NOT
DISASSEMBLE THIS UNIT. DISASSEMBLY OR ATTEMPTED REPAIRS IF ACCOMPLISHED
INCORRECTLY CAN CREATE ELECTRICAL SHOCK HAZARD. ONLY QUALIFIED
TECHNICIANS SHOULD PERFORM SERVICING OF THIS SUCTION UNIT.
!
DANGER
The Heavy Duty Suction Pump is a vacuum suction device designed for the collection of nonflam-
mable fluid materials in medical applications only. Improper use during medical applications can
cause injury or death.
For all medical applications:
1. All suctioning should be done in strict accordance with appropriate procedures that have been
established by a licensed medical authority.
2. Some attachments or accessories may not fit the tubing supplied. All attachments or accesso-
ries should be checked prior to use to assure proper fit.

2. AC POWER SOURCE

CAUTION
!
Item # 18600 is used with a power supply allowing operation on AC voltage (115 VAC, 60 Hz).
However the correct power source must be used to connect to adaptable wall power.

3. ELECTRICAL

Compressor
If the overload protector shuts off the motor frequently, you may have a low voltage situation.
Low voltage can also be suspected when:
1. The motor does not get up to full power or speed.
2. Other motor operated appliances fail to operate properly. Too many motor operated appli-
ances on same circuit.
NOTE
!
The motor of this compressor has a thermal overload protector. If the motor overheats, the over-
load protector will shut off the motor. If this occurs, turn the ON/OFF switch to the OFF position
and allow the motor to cool (approximately 5 minutes). Restart compressor. If the compressor
fails to start, wait an additional 5 minutes before attempting to restart. If this fails, inspect the
fuses on your suction unit. If the unit still does not restart, call your medical equipment provider
immediately.