maintenance

!
NOTE
Use only Drive filters. Do not substitute any other material for this bacteria filter. Improper
substitution may lead to contamination or poor performance;
Fuse Change:
1. Use only the same type of fuse to replace the broken fuse.
2. Fuse specification: Miniature Glass Fuse 5x20mm, Fast Blow (Quick Acting), F2A/250V.
WARNING
!
When this unit or accessories reach the end of their useful lives (depends on using conditions),
residues in collection bottle, tubing and filter may cause respiratory infection. Dump the wasting
collection bottle, tubing and filter into a specific waste bin labeled with medical waste.

troubleshooting

Your Heavy Duty Suction Pump contains no user-serviceable parts. If you believe your unit is
not working properly, BEFORE YOU RETURN IT TO THE MEDICAL EQUIPMENT PROVIDER
WHERE YOU PURCHASED IT, please take a few moments to check for these possible causes:
PROBLEM
Unit does not turn on
Motor runs, no vacuum
Low Vacuum
ACTION
1. Check power source and connection
2. Ensure that the wall outlet is live
3. Check power cord is not damage
1. Verify tubing connection security
2. Check for leaks or tubing kinks
3. Ensure that float shut-off is not activated
4. Check for bottle leaks and cracks
1. Use vacuum adjust knob to increase vacuum level
2. Check system for leaks
3. Adjust vacuum adjust knob and release